weighing indicator error
What is a scale error? /weighing indicator error/What does error mean on a scale?
If the electronic weighing indicator error displays Err H/Err L.
Description: Power-on zero point is too high (higher/lower than maximum capacity + 10%), displayed only when set to certified specification .
Cause: No sensor connected, sensor wire broken, sensor zero point offset.
Solution: Check sensor wiring, re-calibrate.
If the weighing indicator error displays Err N.
Description: Unstable internal value.
Cause: No sensor connected, sensor wire broken, vibration source in the use environment.
Solution: Check sensor wiring, use in a stable environment.
If the electronic scale displays Err 4.
Description: EEPROM read failure.
Cause: EEPROM damaged.
Solution: Replace EEPROM.
If the electronic scale displays Err 10.
Description: Calibration failure.
Cause: No weight added during calibration, abnormal AD data reading.
Solution: Add weight during calibration, check .
If you have difficulty calibrating, please refer to the user manual or contact us.